SCPI Programming Reference
[SENSe:]VOLTage:{AC|DC}:RANGe:AUTO {OFF|ON|ONCE}
[SENSe:]VOLTage:{AC|DC}:RANGe:AUTO?
Disables or enables autoranging for AC and DC voltage measurements and for DC ratio measurements.
Autoranging is convenient because it automatically selects the range for each measurement based on the
input signal.
Specifying ONCE performs an immediate autorange and then turns autoranging off.
This parameter setting is not shared between AC and DC measurements. The parameters are independent for
AC and DC measurements.
The maximum range parameter (MAX) is 1000 V. However, the SAFETY LIMIT on the
front and rear HI/LO input terminals is 750 VAC (rms). The rms voltage is waveform
dependent. A sine wave is limited to 750 VAC (rms), but a 1000 Vpk square wave is
safe. Connection to AC mains power is further limited to CAT II (300 V). See
and Regulatory Information
operation of this instrument.
Parameter
{ON|1|OFF|0}. Default: ON.
Configure DC voltage measurements and perform an immediate autorange. Make and read two meas-
urements:
CONF:VOLT:AC
VOLT:AC:RANG:AUTO ONCE
SAMP:COUN 2
READ?
Typical Response: +1.04530000E+01,+1.04570000E+01
Autoranging goes down a range at less than 10% of range and up a range at greater than 120% of
l
range.
With autoranging enabled, the instrument selects the range based on the input signal.
l
Selecting a fixed range ([SENSe:]<function>:RANGe) disables autoranging.
l
This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset
l
(SYSTem:PRESet).
378
for more information on the safety features and the safe
0 (OFF) or 1 (ON)
Keysight Truevolt Series Operating and Service Guide