This command is nearly identical to *RST. The difference is that *RST resets the instrument for SCPI oper-
ation, and SYSTem:PRESet resets the instrument for front panel operation. As a result, *RST turns the his-
togram and statistics off, and SYSTem:PRESet turns them on.
SYSTem:SECurity:COUNt?
Returns the instrument's security count.
Parameter
Typical Return
(none)
+22
Return the instrument's security count:
SYSTem:SECurity:COUNt?
The security count is incremented by 1 every time you perform an action, other than calibration, that
l
requires the instrument to be unsecured. This includes enabling or disabling interfaces, sanitizing
memory per the NISPOM standard (SYSTem:SECurity:IMMediate), updating firmware, changing the
security code, or overriding the calibration security.
The act of securing or unsecuring the instrument does not increment the counter.
l
Be sure to read the security count upon receiving your instrument.
l
See Also
CALibration:COUNt?
Keysight Truevolt Series Operating and Service Guide
SCPI Programming Reference
401