Performance Verification
Input
Vrms
Frequency
1 kHz
100 mV
50 kHz
300 kHz
1 kHz
1 V
50 kHz
300 kHz
0.03 V
1 kHz
1 V
1 kHz
10 Hz
100 Hz
20 kHz
10 V
50 kHz
100 kHz
300 kHz
1 kHz
100 V
50 kHz
70 V
300 kHz
750 V
1 kHz
210 V
50 kHz
70 V
300 kHz
Q:
Quick performance verification test point.
518
ACV Values for 34461A DMM
Quick
Range
Filter
Check
100 mV 200 Hz
Q
1 V
200 Hz
200 Hz
200 Hz
3 Hz
20 Hz
10 V
200 Hz
Q
200 Hz
Q
100 V
200 Hz
750 V
200 Hz
Error from Nominal
24 hours 90 days
1 year
± 60 µV
± 80 µV
± 90 µV
± 140 µV ± 160 µV ± 170 µV ± 180 µV
± 4.5 mV
± 600 µV ± 800 µV ± 900 µV
± 1.4 mV ± 1.6 mV ± 1.7 mV ± 1.8 mV
± 45 mV
± 2 mV
± 3 mV
± 2.4 mV ± 3.5 mV ± 3.6 mV ± 3.7 mV
± 6 mV
± 8 mV
± 9 mV
± 6 mV
± 8 mV
± 9 mV
± 6 mV
± 8 mV
± 9 mV
± 14 mV
± 16 mV
± 17 mV
± 63 mV
± 68 mV
± 450 mV
± 60 mV
± 80 mV
± 90 mV ± 100 mV
± 140 mV ± 160 mV ± 170 mV ± 180 mV
± 3.3 V
± 450 mV ± 600 mV ± 675 mV ± 750 mV
± 510 mV ± 606 mV ± 627 mV ± 648 mV
± 6.6 V
Keysight Truevolt Series Operating and Service Guide