Les langues disponibles

SCPI Programming Reference
CALCulate:TCHart[:STATe] {ON|1|OFF|0}
CALCulate:TCHart[:STATe]?
Enables or disables trend chart data when controlling the instrument from remote.
Parameter
{OFF|ON|0|1}. Default: OFF.
Enable trend chart data on 500 measurements:
CALCulate:TCHart:STATe ON
CONF:VOLT:DC 10,0.003
SAMP:COUN 500
INIT
FETC?
Typical Response: +9.99383828200000E+00,+9.99352315400000E+00,... (500 measurements)
You must enable this (ON) before sending
l
instrument from remote. If it is disabled or enabled after
ults in a chart with no data.
Trend chart data must be disabled (OFF) to make measurements at the highest speed (50 kHz).
l
This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset
l
(SYSTem:PRESet).
236
Typical Return
OFF or ON
INIT
or
READ?
to see trend chart data when controlling the
INIT
or READ?, viewing the trend chart res-
Keysight Truevolt Series Operating and Service Guide