Télécharger Imprimer la page

Keysight Truevolt Serie Mode D'emploi page 519

Masquer les pouces Voir aussi pour Truevolt Serie:

Publicité

Les langues disponibles

Les langues disponibles

Performance Verification
Input
Vrms
Frequency
1 kHz
100 mV
50 kHz
300 kHz
1 kHz
1 V
50 kHz
300 kHz
0.03 V
1 kHz
1 V
1 kHz
10 Hz
100 Hz
20 kHz
10 V
50 kHz
100 kHz
300 kHz
1 kHz
100 V
50 kHz
70 V
300 kHz
750 V
1 kHz
210 V
50 kHz
70 V
300 kHz
Q:
Quick performance verification test point.
520
ACV Values for 34465A and 34470A DMMs
Quick
Range
Filter
Check
100 mV 200 Hz
Q
1 V
200 Hz
200 Hz
200 Hz
3 Hz
20 Hz
10 V
200 Hz
Q
200 Hz
Q
100 V
200 Hz
750 V
200 Hz
Error from Nominal
24 hours 90 days
1 year
± 40 µV
± 60 µV
± 70 µV
± 80 µV
± 90 µV
± 100 µV ± 110 µV
± 1.1 mV
± 400 µV ± 600 µV ± 700 µV ± 800 µV
± 800 µV ± 900 µV
± 1 mV
± 11 mV
± 2 mV
± 2.2 mV ± 2.4 mV ± 2.5 mV ± 2.6 mV
± 4 mV
± 6 mV
± 7 mV
± 8 mV
± 9 mV
± 10 mV
± 20 mV
± 110 mV
± 40 mV
± 60 mV
± 70 mV
± 80 mV
± 90 mV ± 100 mV ± 110 mV
± 800 mV
± 750 mV ± 900 mV ± 975 mV
± 330 mV ± 351 mV ± 372 mV ± 393 mV
± 1.5 V
Keysight Truevolt Series Operating and Service Guide
2 years
± 80 µV
± 1.1 mV
± 8 mV
± 11 mV
± 80 mV
± 1.1 V

Publicité

loading

Ce manuel est également adapté pour:

Truevolt 34460aTruevolt 34461aTruevolt 34465aTruevolt 34470a