SCPI Programming Reference
MEASure:TEMPerature? [{FRTD|RTD|FTHermistor|THERmistor|TCouple|DEFault} [, {<type>|DEFault} [,1 [,
{<resolution>|MIN|MAX|DEF}]]]]
Sets all measurement parameters and trigger parameters to their
urements and immediately triggers a measurement. The results are sent directly to the instrument's out-
put buffer.
<probe_type>: {FRTD|RTD|FTHermistor|THERmistor|TCouple}. Default: FRTD.
<type>: 85 (only possible value for RTD/FRTD), 5000 (only possible value for THERmis-
tor/FTHermistor), or E, J, K, N, R, T (TCouple).
Resolution Table
<resolution>: See
default is equivalent to 10 PLC.
Configure a 4-wire RTD measurement with default resolution. Then make and read one measurement:
MEAS:TEMP? FRTD,85
Typical Response:+2.12320000E+01
Thermocouple measurements (TC) apply only to the 34465A and 34470A.
l
For temperature measurements, the instrument internally selects the range; you cannot select the
l
range to be used. The range ("1" is the only allowed value) is used only in conjunction with resolution
to determine NPLC.
The <resolution>parameter only determines the integration time; it does not specify the resolution of
l
the temperature measurement. This parameter is optional; however, if you specify <resolution>, you
must also specify "1" as the implied range parameter. For example: CONF:TEMP
RTD,85,1,0.000001 selects the 10 PLC integration time on the 34461A.
To change temperature units, use UNIT:TEMPerature.
l
For RTD and thermistor measurements, the instrument autoranges to the correct range for the trans-
l
ducer resistance measurement. For thermocouple measurements, the 100 mV range is selected.
For thermocouple measurements, the internal reference is selected. See
l
ature:TCouple:RJUNction:TYPE.
If the input signal is greater than can be measured on the specified manual range, the instrument dis-
l
plays the word Overload on front panel and returns "9.9E37" from the remote interface.
288
Parameter
Range, Resolution and NPLC.
or
default values
for temperature meas-
Typical Return
(none)
The
SENSe:TEMPer-
Keysight Truevolt Series Operating and Service Guide