AC Voltage Low Frequency Gain and Flatness Calibration
This calibration procedure applies to all models (34460A/61A/65A/70A).
Review
Test Considerations
Configuration: AC Voltage
1. Configure the range shown in the table below.
2. Apply the input signal shown in the Input column.
3. Enter the actual applied input voltage amplitude (see
Step Succeeded message indicates success; if the display shows Calibration Step Failed, check the
input value, range, function, and entered calibration value and repeat the calibration step.
4. Repeat steps 1 through 3 for each flatness calibration point shown in the table.
5. Store the new calibration constants (see
Each voltage and frequency takes approximately 2.5 seconds to complete.
Input
Vrms
Frequency
7 V
10 Hz
7 V
40 Hz
Keysight Truevolt Series Operating and Service Guide
and
Flatness Calibration Considerations
Storing the Calibration
Instrument
Range
10 V
10 V
Calibration Adjustment Procedures
before performing any calibration.
Entering Calibration
Values). A Calibration
Constants).
Next Procedure