Keysight Truevolt Serie Mode D'emploi page 348

Masquer les pouces Voir aussi pour Truevolt Serie:

Publicité

Les langues disponibles

Les langues disponibles

[SENSe:]{RESistance|FRESistance}:NULL[:STATe] {ON|1|OFF|0}
[SENSe:]{RESistance|FRESistance}:NULL[:STATe]?
Enables or disables the null function for all resistance measurements.
This parameter is common to 2-wire and 4-wire resistance measurements. Using the FRESistance version of
this command or query is identical to using the RESistance version.
Parameter
{ON|1|OFF|0}. Default: OFF.
Configure 2-wire resistance measurements, using the null function to remove 100 mΩ of wiring res-
istance. Make and read two measurements:
CONF:RES
RES:NULL:STAT ON;VAL .1
SAMP:COUN 2
READ?
Typical Response: +1.04530000E+02,+1.04570000E+02
Enabling the scaling function also enables automatic null value selection
l
{RESistance|FRESistance}:NULL:VALue:AUTO
To set a fixed null value, use: [SENSe:]{RESistance|FRESistance}:NULL:VALue.
l
The instrument disables the null function after a Factory Reset (*RST), Instrument Preset
l
(SYSTem:PRESet), or CONFigure function.
Keysight Truevolt Series Operating and Service Guide
Typical Return
0 (OFF) or 1 (ON)
ON).
SCPI Programming Reference
(SENSe:
349