Calibration Adjustment Procedures
AC Voltage Gain and Flatness Calibration
This calibration procedure applies to all models (34460A/61A/65A/70A).
Review
Test Considerations
Configuration: AC Voltage
All AC calibration procedures use the 3 Hz bandwidth measurement filter.
1. Configure each function and range in the order shown in the table below.
2. Apply the input signal shown in the Input column.
3. Enter the actual applied input voltage amplitude (see
Step Succeeded message indicates success; if the display shows Calibration Step Failed, check the
input value, range, function, and entered calibration value and repeat the calibration step.
4. Repeat steps 1 through 3 for each flatness calibration point shown in the table.
5. Store the new calibration constants (see
Each voltage and frequency takes approximately 2.5 seconds to complete.
Vrms
5 kHz, 10 kHz, 20 kHz, 35 kHz, 50 kHz, 75 kHz, 100 kHz,
100 mV
200 kHz, 300 kHz, 390 kHz, 400 kHz, 220 Hz, 1000 Hz
5 kHz, 10 kHz, 20 kHz, 35 kHz, 50 kHz, 75 kHz, 100 kHz,
1 v
200 kHz, 300 kHz, 390 kHz, 400 kHz, 220 Hz, 1000 Hz
5 kHz, 10 kHz, 20 kHz, 35 kHz, 50 kHz, 75 kHz, 100 kHz,
10 V
200 kHz, 300 kHz, 390 kHz, 400 kHz, 220 Hz, 1000 Hz
5 kHz, 10 kHz, 20 kHz, 35 kHz, 50 kHz, 75 kHz, 100 kHz,
50 V
200 kHz, 300 kHz, 390 kHz, 400 kHz, 220 Hz, 1000 Hz
500 V
560
and
Flatness Calibration Considerations
Storing the Calibration
Input
Frequency
1000 Hz
before performing any calibration.
Entering Calibration
Values). A Calibration
Constants).
Instrument
Range
100 mV
1 V
10 V
100 V
750 V
Next Procedure
Keysight Truevolt Series Operating and Service Guide