SCPI Programming Reference
width.
Use
READ?
or
INITiate
l
See Also
CONFigure?
MEASure:CURRent:{AC|DC}?
[SENSe:]CURRent:AC:BANDwidth
CONFigure:DIODe
Sets all measurement parameters and trigger parameters to their
Parameter
(none)
Configure, make, and read a default diode measurement:
CONF:DIOD
READ?
Typical Response: +1.32130000E-01
The range and resolution are fixed for diode tests: the range is 1 VDC (with a 1 mA current source out-
l
put).
The voltage is displayed on the front panel if it is between 0 and 5.05 V. The instrument beeps when
l
the signal transitions into the range between 0.3 and 0.8 V (unless beeper is disabled). If the signal
exceeds 5.05 V, the front panel shows "OPEN", and the value returned from SCPI is 9.9E37.
The FETCh?, READ?, and
l
value.
Use
READ?
or
INITiate
l
248
to start the measurement.
Typical Return
(none)
MEASure:DIODe?
queries return the measured voltage, regardless of its
to start the measurement.
default values
for diode tests.
Keysight Truevolt Series Operating and Service Guide