Features and Functions
Data Log and Digitize Mode Default Settings
On entering data log or digitize mode, the DMM configures these settings:
Trigger count set to 1 (trigger count is normally infinite when in Local and is not settable from the
l
front panel).
Secondary measurements are turned off.
l
Math smoothing is turned off.
l
Statistics are cleared.
l
Histogram is cleared.
l
Trend chart is changed from the continuous, bucketized mode, to simple data graph.
l
Additional Data Log Default Settings
On entering data log mode, the DMM configures these settings:
Trigger source is set to auto.
l
Trigger delay is set to auto.
l
Pretrigger count is set to zero.
l
Samples per trigger is set according to the data log duration (time or samples).
l
Sample timer is put in timer (not immediate) mode and sample time is set according to the data log
l
sample interval.
48
Keysight Truevolt Series Operating and Service Guide