MEASure:{RESistance|FRESistance}? [{<range>|AUTO|MIN|MAX|DEF} [, {<resolution>|MIN|MAX|DEF}]]
Sets all measurement and trigger parameters to their
or 2-wire (RESistance) measurements, and immediately triggers a measurement. The results are sent dir-
ectly to the instrument's output buffer.
<range>: 100 Ω, 1 kΩ, 10 kΩ, 100 kΩ, 1 MΩ, 10 MΩ, 100 MΩ, 1
GΩ, AUTO,or DEFault. Default: AUTO.
Resolution Table
<resolution>: See
NPLC.
The default is equivalent to 10 PLC.
Specify <resolution> in the measurement's units (V, A, Hz, Ω,
and so on).
Configure 4-wire resistance measurements using the 100 Ω range with default resolution. Then make and
read one measurement:
MEAS:FRES? 100
Typical Response:+8.54530000E+01
The 1 GΩ range applies only to the 34465A and 34470A.
l
You can let autoranging select the measurement range, or you can manually select a fixed range.
l
Autoranging conveniently selects the range for each measurement based on the input signal. For fast-
est measurements, use manual ranging (autoranging may require additional time for range selection).
Autoranging (AUTO or DEFault), will generate an error if you specify a <resolution> because the instru-
l
ment cannot accurately resolve the integration time (especially if the input continuously changes). If
your application requires autoranging, specify DEFault for the <resolution> or omit the <resolution>
altogether.
Autoranging goes down a range at less than 10% of range and up a range at greater than 120% of
l
range.
If the input signal is greater than can be measured on the specified manual range, the instrument dis-
l
plays the word Overload on front panel and returns "9.9E37" from the remote interface.
Keysight Truevolt Series Operating and Service Guide
Parameter
Range, Resolution and
or
SCPI Programming Reference
default values
for 4-wire resistance (FRESistance)
Typical Return
(none)
287