[SENSe:]VOLTage:{AC|DC}:NULL:VALue {<value>|MIN|MAX|DEF}
[SENSe:]VOLTage:{AC|DC}:NULL:VALue? [{MIN|MAX|DEF}]
Stores a null value for voltage measurements.
This parameter setting is not shared between AC and DC measurements. The parameters are independent for
AC and DC measurements.
Parameter
-1200 to +1200 V. Default: 0.
Configure AC voltage measurements, using the null function to subtract 100 mV from the measurements.
Make and read two measurements:
CONF:VOLT:AC
VOLT:AC:NULL:STAT ON;VAL 100 mV
SAMP:COUN 2
READ?
Typical Response: +1.04530000E+00,+1.04570000E+00
Specifying a null value disables automatic null value selection
l
{AC|DC}:NULL:VALue:AUTO
To use the null value, the null state must be on
l
The null function is not available for DC ratio measurements.
l
This parameter is set to its default value after a Factory Reset (*RST), Instrument Preset
l
(SYSTem:PRESet), or CONFigure function.
Keysight Truevolt Series Operating and Service Guide
+1.00000000E-02
OFF).
([SENSe:]VOLTage:{AC|DC}:NULL:STATe
SCPI Programming Reference
Typical Return
([SENSe:]VOLTage:
ON).