DC Current Gain Calibration
This calibration procedure applies to the 34460A and 34461A only.
Review
Test Considerations
Configuration: DC Current
1. Configure each function and range in the order shown in the table below.
2. Apply the input signal shown in the Input column.
3. Enter the actual applied input current as the Cal Value (see
Step Succeeded message indicates success; if the display shows Calibration Step Failed, check the
input value, range, function, and entered calibration value and repeat the calibration step.
4. Repeat steps 1 through 3 for each gain calibration point shown in the table.
5. Store the new calibration constants (see
Each range takes less than 4 seconds to complete.
Instrument Settings
Input
Function
100 µA
1 mA
10 mA
DC Current
100 mA
1 A
Keysight Truevolt Series Operating and Service Guide
and
Gain Calibration Considerations
Storing the Calibration
Range
100 µA
1 mA
10 mA
100 mA
1 A
Calibration Adjustment Procedures
before beginning this procedure.
Entering Calibration
Values).A Calibration
Constants).
Next Procedure
551