SCPI Programming Reference
CALCulate:AVERage:ALL?
CALCulate:AVERage:AVERage?
CALCulate:AVERage:COUNt?
CALCulate:AVERage:MAXimum?
CALCulate:AVERage:MINimum?
CALCulate:AVERage:PTPeak?
CALCulate:AVERage:SDEViation?
The CALCulate:AVERage:ALL? query returns the arithmetic mean (average), standard deviation, minimum
value and maximum value of all measurements taken since the statistics were last cleared. The count and
peak-to-peak statistics are not returned by CALCulate:AVERage:ALL?.
The other six queries listed above return individual values.
Parameter
(none)
Return the statistics of 100 frequency measurements:
CONF:FREQ 1.0E3,.001
SAMP:COUN 100
CALC:AVER:STAT ON
INIT
*WAI
CALC:AVER:ALL?
Typical Response: +1.00520000E+03,+1.00512000E+03,+1.00527000E+03,+4.13500000E-01