Calibration Adjustment Procedures
AC Current 10 A Gain Calibration
This calibration procedure applies to the 34461A, 34465A, and 34470A only (not applicable to the
34460A).
Review
Test Considerations
Configuration: AC Current
All AC calibration procedures use the 3 Hz bandwidth measurement filter.
1. Select the 10 A range.
2. Apply the 5 A input signal at 1000 Hz.
3. Enter the actual applied input current as the Cal Value (see
Step Succeeded message indicates success; if the display shows Calibration Step Failed, check the
input value, range, function, and entered calibration value and repeat the calibration step.
4. Store the new calibration constants (see
This calibration takes approximately 2.5 seconds to complete.