-221,"Settings conflict; pretrigger count limited to 10,000 with CALC enabled; SAMP:COUN:PRET chan-
ged"
-221,"Settings conflict; pretrigger not supported on current measurement function; SAMP:COUN:PRET 0
set"
-221,"Settings conflict; pretrigger SAMP:COUN:PRETrig >= SAMP:COUN; SAMP:COUN:PRET changed"
-221,"Settings conflict; SAMP:COUN limited to memory size in pretrigger mode; SAMP:COUN changed"
-221,"Settings conflict; scaling not allowed in current measurement function; CALC:SCAL:STAT OFF set"
-221,"Settings conflict; smoothing not allowed in current measurement function; CALC:SMO:STAT OFF
set"
-221,"Settings conflict; statistics not allowed in current measurement function; CALC:AVER:STAT OFF
set"
-221,"Settings conflict; TRIG:LEVel value < minimum for current measurement function/range; TRIG:LEV
MIN selected"
-221,"Settings conflict; TRIG:LEVel value > maximum for current measurement function/range; TRIG:LEV
MAX selected"
-221,"Settings conflict; trigger count must be 1 in pretrigger mode; TRIG:COUN 1 set"
-221,"Settings conflict; trigger source is BUS"
-222,"Data out of range"
-222,"Data out of range; value clipped to lower limit"
-222,"Data out of range; value clipped to upper limit"
-222,"Data out of range; value requires license DIG"
-222,"Data out of range; value requires license MEM"
-223,"Too much data"
-224,"Illegal parameter value"
-225,"Out of memory; measurement data overrun"
-230,"Data corrupt or stale"
-231,"Data questionable; requested sample timing not met"
-240,"Hardware error; cannot communicate with measurement hardware"
-240,"Hardware error; CPU board initialization failed"
-240,"Hardware error; GPIB interface failed"
-240,"Hardware error; measurement FPGA FIFO overflow"
-240,"Hardware error; measurement hardware initialization failed"
-241,"Hardware missing"
-250,"Mass storage error: file read/write error"
488