Features and Functions
1. Select the measurement function and make connections to the DUT (see
2. Select Data Log mode (press [Acquire] > Acquire > Data Log).
3. Specify the sample interval (time between readings) for example, 20 ms.
4. Specify the duration as amount of time or number of readings.
5. Specify when to start data logging (delay or time of day). You can only use auto trigger (default) or single
trigger (press [Single])for data logging.
6. Select whether to log data to memory or internal or external data file(s).
7. Press [Run/Stop] or [Single]. Data logging starts when the specified delay has elapsed or the time of day
occurs (specified in Step 5). Data logging will stop after the specified duration (time or number of read-
ings) has occurred or after you press [Run/Stop] again.
Detailed Data Logging Steps
Note: For more information, on any of the softkeys described below, such as the range of values for a par-
ticular setting, press and hold the softkey to display help for that softkey.
Step 1: Select the measurement function and make connections to the DUT (see
details). For example, press DCV and configure the test leads as shown.
Step 2: Press [Acquire] on the front panel to display these softkeys:
Press the Acquire softkey.
Press the Data Log softkey. The Data Log menu opens:
Step 3: Press Sample Interval and specify the time interval between samples (readings).
Note: You may see this message when configuring data logging: Sample interval is limited by meas-
urement settings. The measurement time is determined by the measurement function, NPLC, Aperture,
82
Measurements
Measurements
Keysight Truevolt Series Operating and Service Guide
for details).