Use the following procedure to configure the line voltage selector:
Step 1
Lift tab (1) and pull the fuse drawer (2) from the rear panel.
Step 2
Remove the line-voltage selector and rotate it so the correct voltage will
appear in the fuse holder window.
Step 3
Verify that the correct fuse is installed. To replace a blown fuse
or verify the correct fuse,
pull it gently from the fuse drawer and insert the correct work-
ing fuse. Use only a 5x20 mm,
time-lag, 0.25 A, 250 V certified fuse. The Keysight part num-
ber is 2110-0817.
Step 4
Replace the fuse holder assembly by sliding it into the rear panel.
Product Grounding
The instrument is a Class 1 product and is provided with a grounding-type power cord set. The instrument
chassis and cover are connected to the instrument electrical ground to minimize shock hazard. The ground
pin of the cord set plug must be firmly connected to the electrical ground (safety ground) terminal at the
power outlet. Any interruption of the protective earth (grounding) conductor or disconnection of the pro-
tective earth terminal will cause a potential shock hazard that could result in personal injury or death.