Télécharger Imprimer la page

Keysight Truevolt Serie Mode D'emploi page 353

Masquer les pouces Voir aussi pour Truevolt Serie:

Publicité

Les langues disponibles

Les langues disponibles

SCPI Programming Reference
See Also
[SENSe:]{RESistance|FRESistance}:RANGe:AUTO
[SENSe:]{RESistance|FRESistance}:RANGe:AUTO {OFF|ON|ONCE}
[SENSe:]{RESistance|FRESistance}:RANGe:AUTO?
Disables or enables autoranging for all resistance measurements. Autoranging is convenient because it
automatically selects the range for each measurement based on the input signal. Specifying ONCE per-
forms an immediate autorange and then turns autoranging off.
This parameter is common to 2-wire and 4-wire resistance measurements. Using the FRESistance version of
this command or query is identical to using the RESistance version.
Parameter
{ON|1|OFF|0}. Default: ON.
Configure 2-wire resistance measurements and perform an immediate autorange. Make and read two
measurements:
CONF:RES
RES:RANG:AUTO ONCE
SAMP:COUN 2
READ?
Typical Response: +1.04530000E+03,+1.04570000E+03
Autoranging goes down a range at less than 10% of range and up a range at greater than 120% of
l
range.
With autoranging enabled, the instrument selects the range based on the input signal.
l
Selecting a fixed range ([SENSe:]<function>:RANGe) disables autoranging.
l
This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset
l
(SYSTem:PRESet).
354
Typical Return
0 (OFF) or 1 (ON)
Keysight Truevolt Series Operating and Service Guide

Publicité

loading

Ce manuel est également adapté pour:

Truevolt 34460aTruevolt 34461aTruevolt 34465aTruevolt 34470a