Features and Functions
Continuity
This section describes how to configure continuity tests from the front panel.
Step 1: Configure the test leads as shown.
Step 2: Press [Cont] on the front panel to open a menu where you can enable or disable the beeper for all
functions that use the beeper (limits, probe hold, diode, continuity, and errors).
Continuity measurements behave as follows:
≤ 10 Ω displays measured resistance and beeps (if beeper enabled)
10 Ω to 1.2 kΩ displays measured resistance without beeping
> 1.2 kΩ displays OPEN with no beep