6.
Rotate the dilator 90 degrees to unlock it from the introducer sheath (Figure 13). Being careful to leave the
introducer sheath in place, remove the directional guide and dilator (Figure 14).
7.
The tined lead is provided with the straight stylet. If desired, remove the straight stylet from the lead and
replace with the curved stylet.
8.
Carefully insert the tined lead (with stylet) into the introducer sheath until visual marker B on the tined lead is
Figure 14: Remove the directional guide and dilator
aligned with the top of the introducer sheath handle. Confirm with fluoroscopy that all electrodes on the lead
are proximal to the distal tip of the sheath (Figure 15).
Notes:
• Use the tined lead markers to determine at what point the tines will be deployed. Refer to Figure 1 on page 16.
• The stylet supplied is 5mm longer than the tined lead and should not be forced into the lead.
• Sterile water can be used as a lubricant if necessary to insert the tined lead into the introducer sheath.
PROPOSED AXONICS FIGURE
PROPOSED AXONICS FIGURE
Figure 13. Unlock dilator from introducer sheath.
Figure 13: Unlock dilator from introducer
Figure 14. Remove the directional guide and dilator.
sheath
18