60 ALLOCATE REAL Wave_form(1:Num_samples)!CREE UN TABLEAU POUR
65
70 DATA ALLOCATE REAL Samp(1:Num_samples)!CREE UN TABLEAU POUR LES
75
80 ASSIGN @Dvm TO 722
90 ASSIGN @Int_samp TO BUFFER Int_samp(*)!AFFECTATION DU NOM
95
100 OUTPUT @Dvm;"PRESET FAST;LEVEL;SLOPE;SSRC LEVEL;SSDC 10"
102 !MESURES RAPIDES, TARM SYN, SOURCE SYNC LEVEL 0V, PENTE
105 !POSITIVE (VALEURS PAR DEFAUT), SOUS-ECHANTILLONNAGE (FORMAT
107 !DE SORTIE SINT), GAMME 10 V
110 OUTPUT @Dvm;"SWEEP ";Eff_int,Num_samples !INTERVALLE REEL DE
115
120 TRANSFER @Dvm TO @Int_samp;WAIT !EVENEMENT SYNCHRONE,
121 !TRANSFERT DES LECTURES DANS LE TABLEAU DES ENTIERS DE
122 !L'ORDINATEUR; PAS DE CONVERSION DE DONNEES REQUISES PUISQUE
125 !LE FORMAT INTEGER DE L'ORDINATEUR EST IDENTIQUE A SINT (MAIS
127 !TABLEAUX DES ENTIERS REQUIS)
130 OUTPUT @Dvm;"ISCALE?"
135
140 ENTER @Dvm; S
150 OUTPUT @Dvm;"SSPARM?"
155
160 ENTER @Dvm;N1,N2,N3
165
170 FOR I=1 TO Num_samples
180 Samp(I)=Int_samp(I)
185 !ENTIER AU FORMAT REELCOBLIGATOIRE POUR EVITER UN
187 !DEPASSEMENT ENTIER SUR LA LIGNE SUIVANTE)
190 R=ABS(Samp(I))
195
200 IF R>=32767 THEN PRINT "OVLD" !SI SURCHARGE, IMPRESSION D'UN
205
210 Samp(I)=Samp(I)*S
215
220 Samp(I)=DROUND(Samp(I),4)!VALEUR ARRONDIE A 4 CHIFFRES
Keysight 3458A Guide d'utilisation
!MEMOIRE-TAMPON
!LES DONNEES TRIEES
!ECHANTILLONS
!AFFECTATION DE L'ADRESSE DU MULTIMETRE
!D'ACCES A LA MEMOIRE-TAMPON
!2 µs, 1000 ECHANTILLONS
!INTERROGATION DU FACTEUR D'ECHELLE POUR
!LE FORMAT SINT
!ENTRE LE FACTEUR D'ECHELLE
!INTERROGATION DES PARAMETRES DE SOUS
!ECHANTILLONNAGE
!ENTRE LES PARAMETRES DE SOUS
!ECHANTILLONNAGE
!CONVERSION DE CHAQUE LECTURE DU FORMAT
!UTILISATION DE LA VALEUR ABSOLUE POUR
!VERIFIER SURCHARGE
!MESSAGE
!MULTIPLICATION DE LA LECTURE PAR LE
!FACTEUR D'ECHELLE