Télécharger Imprimer la page

Keysight Technologies 3458A Guide D'utilisation page 732

Multimètre

Publicité

5
Numérisation
100 OUTPUT @Dvm;"LEVEL 250,DC" DECLENCHEMENT A 250% DE LA GAMME (25 V)
110 OUTPUT @Dvm; "TRIG LEVEL" !EVENEMENT DE DECLENCHEMENT FAR NIVEAU
120 OUTPUT @Dvm; "MEM FIFO"
130 TRANSFER @Dvm TO @Buffer;WAIT !TRANSMET LES ECHANTILLONS AU
135
140 OUTPUT @Dvm; "ISCALE'?"
145
150 ENTER @Dvm;S
160 FOR I=l TO Num_samples
170 ENTER @Buffer USING "#, W, W" ;J,K !ENTRE UN MOT DE 16 BITS
172 !NOTATION COMPLEMENT A 2 DANS CHAQUE VARIABLE J ET K ( # =
175 !FIN D'INSTRUCTION NON REQUISE; W = ENTRER DONNEES SOUS
177 !FORME D'ENTIERS 16 BITS NOTATION COMPLEMENT A 2)
180 Samp(I)=(J*65536.+K+65536.*(K<O)) !CONVERSION EN NOMBRE REEL
190 R=ABS(Samp(I))
195
200 IF R>=2147483647 THEN PRINT "OVLD" !SI SURCHARGE, IMPRESSION
205
210 Samp(I)=Samp(I)*S
220 Samp(I)=DROUND(Samp(I),8) !ARRONDIT LES LECTURES CONVERTIES
230 PRINT Samp(I)
240 NEXT I
250 END
228
!VALIDE LA MEMOIRE DE LECTURE, MODE FIFO
!CONTROLEUR
!INTERROGATION DU FACTEUR D'ECHELLE FOUR
!LE FORMAT DINT
!ENTRE LE FACTEUR D'ECHELLE
!UTILISATION DE LA VALEUR ABSOLUE FOUR
!VERIFIER LA SURCHARGE
!D'UN MESSAGE
!APPLICATION DU FACTEUR D'ECHELLE
!IMPRESSION DES LECTURES
Keysight 3458A Guide d'utilisation

Publicité

loading