Example
OUTPUT 722; "APER 10E-6" !SETS APERTURE TO 10 MICROSECONDS
ARANGE
Autorange. Enables or disables the autorange function.
Syntax
ARANGE [control]
control
The control parameter choices are:
control
parameter
OFF
ON
ONCE
Power-on control = ON.
Default control = ON.
Remarks
– With autorange enabled, the multimeter samples the input signal before each
– Refer to the
– Autorange does not operate for direct- or sub-sampled measurements (DSAC,
– Query command: The ARANGE? query command returns a response
Keysight 3458A User's Guide
Numeric
query equiv.
0
1
2
reading and selects the appropriate range.
FUNC
or
measurement function.
DSDC, SSAC, or SSDC command) or when using the TIMER sample event or
the SWEEP command.
indicating the present autorange mode. Refer to
front of this chapter for more information.
Description
Disables autorange algorithm
Enables autorange algorithm
Causes the multimeter to autorange once, then disables
autoranging
RANGE
command for a listing of the ranges for each
Command Reference
Query commands
near the
6
239