Keysight Technologies 3458A Guide D'utilisation page 237

Multimètre

Publicité

ACDCI, ACDCV, ACI, ACV
Refer to the
ADDRESS
Sets the multimeter's GPIB address (from the front panel only). The address is
stored in continuous memory and is not lost when power is removed.
Syntax
ADDRESS value
value
The value parameter is an integer from 0 to 31.
Power-on value = previously stored address (factory setting = 22).
Default value = none; parameter required.
Remarks
– When you specify address 31, it doesn't actually change the multimeter's
– The controller's address is typically 21. Do not use the controller's address for
– When the multimeter detects a CMOS RAM failure (auxiliary error bit 12). It
– ADDRESS? query: From the multimeter's front panel, you can read the present
– Related commands: ID?
Keysight 3458A User's Guide
FUNC
command.
address but sets the multimeter to the Talk Only mode. In this mode, the
multimeter outputs readings directly to an GPIB printer without a controller on
the bus (you must use the ASCII output format). The multimeter's TALK
annunciator illuminates when in Talk Only mode. You cannot specify address
31 with a controller on the bus. To remove the multimeter from Talk Only
mode, press the Reset key or specify an address other than 31.
any other device on the GPIB bus.
sets the address to 22.
address using the Address key (shifted Local key).
Command Reference
6
237