Command Reference
– Query command: The RATIO? query command returns the present ratio mode.
– Related commands: ACDCV, ACV, DCV, SETACV
Example
10 OUTPUT 722;"PRESET NORM" !SUSPEND READINGS,NRDGS=1
20 OUTPUT 722;"ACV" !SELECT AC VOLTAGE MEASUREMENTS
30 OUTPUT 722;"SETACV SYNC" !SYNCHRONOUS ACV MEASUREMENTS
40 OUTPUT 722;"RATIO ON" !ENABLE RATIO MEASUREMENTS
50 OUTPUT 722;"TRIG SGL" !TRIGGER MEASUREMENT
60 ENTER 722;A !ENTER RATIO
70 PRINT A !PRINT RATIO
80 END
RES
Resolution. Specifies reading resolution.
Syntax
RES [%_resolution]
%
For frequency and period measurements, the %_resolution parameter specifies
the digits of resolution and the gate time as shown below. (%_resolution also
affects the reading rate. Refer to the
more information.) If you default the %_resolution parameter for frequency or
period measurements, the multimeter uses .00001.
328
voltage range is ±12 VDC (autorange only). To specify ratio measurements,
you first select the measurement function (and the measurement method for
ACV or ACDCV) and then enable ratio measurements with the RATIO
command (see example below).
Refer to
Query commands
_
resolution
near the front of this chapter for more information.
"Appendix A: Specifications"
on page 409 for
Keysight 3458A User's Guide