Weighted
value
1
2
4
8
16
32
64
128
256
512
1024
2048
4096
8192
16384
Power-on value = 32767 (all enabled).
Default value = 32767 (all enabled).
Remarks
– When an error occurs, it sets the corresponding bit in the error register
– Query command: The EMASK? query command returns the weighted sum of
– Related commands: AUXERR?, ERR?, ERRSTR?, RQS, STB?
Keysight 3458A User's Guide
Bit
Error cond itions
number
0
Hardware error (see
1
Calibration error
2
Trigger too fast error
3
Syntax error
4
Command not allowed from remote (ADDRESS command)
5
Undefined parameter received
6
Parameter out of range
7
Memory error
8
Destructive overload detected
9
Out of calibration
10
Calibration required
11
Settings conflict (memory improperly configured for sub-sampling)
12
Math error (divide by 0, integer overflow, etc.)
Subprogram error (calling a deleted sub, CONT with no PAUSE, SUBEND or
13
PAUSE only allowed in sub, SCRATCH, DELSUB, CONT, not allowed in sub)
14
System error
regardless of whether or not it has been enabled by the EMASK command.
Disabling an error bit prevents it from setting the error bit in the status register
only, and thereby generating a service request.
all enabled error conditions (see example below).