6
Command Reference
PRESET
Configures the multimeter to one of three predefined states.
Syntax
PRESET [type]
type
Specifies the NORM, FAST, or DIG preset state (the numeric query equivalents of
these parameters are 1, 0, and 2, respectively).
Power-on type = not applicable.
Default type = NORM.
NORM
PRESET NORM is similar to RESET but optimizes the multimeter for remote
operation. Executing PRESET NORM executes the following commands:
ACBAND 20,2E+6MEM OFF (last memory operation set to FIFO)
AZERO ONMFORMAT SREAL
BEEP ONMMATH OFF
DCV AUTONDIG 6
DELAY -1NPLC 1
DISP ONNRDGS 1,AUTO
FIXEDZ OFF OCOMP OFF
FSOURCE ACVOFORMAT ASCII
INBUF OFFTARM AUTO
LOCK OFFTIMER 1
MATH OFFTRIG SYN
All math registers set to 0 except:
DEGREE = 20
PERC = 1
REF = 1
RES = 50
SCALE = 1
318