SETACV
Set ACV. Selects the RMS conversion technique to be used for AC or AC+DC
voltage measurements.
Syntax
SETACV [type]
type
The type parameter is used to select the measurement method: analog, random
sampling, or synchronous sampling. The parameters are:
type
parameter
ANA
RNDM
SYNC
Power-on type = ANA.
Default type = ANA.
Remarks
– Bandwidth limitations vary with the conversion technique selected. See the
– Query command: The SETACV? query command returns the present AC
– Related commands: ACBAND, ACDCV, ACV, FUNC, SSRC
Example
10 OUTPUT 722; "SETACV SYNC" !SPECIFIES SYNCHRONOUS SAMPLING (DC
COUPLED)
20 OUTPUT 722;"ACDCV" !SELECTS AC+DC VOLTAGE MEASUREMENTS