operation
parameter
RMS
SCALE
STAT
CTHRM2K
CTHRM10K
FTHRM2K
FTHRM10K
CRTD85
CRTD92
FRTD85
FRTD92
Power-on operation_a,operation_b = OFF,OFF.
Default operation_a,operation_b = OFF,OFF.
Keysight 3458A User's Guide
Numeric
Description
equiv.
Result = squares reading, applies FILTER operation, takes square
12
root.
13
Result = (reading-OFFSET register) / SCALE register.
Performs statistical calculations on the present set of readings and
stores results in these registers:
SDEV = standard deviation
14
MEAN = average of readings
NSAMP = number of readings
UPPER = largest reading
LOWER = smallest reading
Result = temperature (Celsius) of a 2 kW thermistor (40653A).
16
Function must be OHM or OHMF.
Result = temperature (Celsius) of a 10 kΩ thermistor (40653C).
17
Function must be OHM or OHMF.
Result = temperature (Fahrenheit) of a 2 kW thermistor (40653A).
18
Function must be OHM or OHMF.
Result = temperature (Fahrenheit) of a 10 kW thermistor (40653C).
19
Function must be OHM or OHMF.
Result = temperature (Celsius) of 100 W RTD with alpha of
20
0.00385 (40654A or 406548). Function must be OHM or OHMF.
Result = temperature (Celsius) of 100 W RTD with alpha of
21
0.003916. Function must be OHM or OHMF.
Result = temperature (Fahrenheit) of 100 W RTD with alpha of
22
0.00385 (40654A or 406548). Function must be OHM or OHMF.
Result = temperature (Fahrenheit) of 100 W RTD with alpha of
23
0.003916. Function must be OHM or OHMF.
Command Reference
6
287