Télécharger Imprimer la page

Keysight Technologies 3458A Guide D'utilisation page 138

Multimètre

Publicité

4
Making Measurements
Table 4-2
Trigger
138
Event combinations (continued)
Trigger
arm
Sample event Description
event
event
SGL
AUTO
AUTO, EXT,
SGL
EXT
TIMER,
LINE, LEVEL
SGL
EXT
AUTO, EXT,
SGL
LEVEL
TIMER,
LEVEL
SGL
LEVEL
SYN, LINE
AUTO, EXT,
SGL
LINE
TIMER,
LINE
SGL
LINE
SYN, LEVEL
SGL
SGL
SGL
SYN
AUTO, EXT,
SGL
SYN
TIMER,
LINE, LEVEL
After executing the TARM SGL command, one reading is taken
per sample event until the specified number of readings are
Any
completed. The trigger arm event then becomes HOLD. When
using the SYN sample event, the input buffer must be enabled or
you must suppress cr lf when sending the TARM SGL command.
After executing the TARM SGL command followed by a negative
edge transition on the Ext Trig input, one reading is taken per
sample event, until the specified number of readings are
completed. The trigger arm, event then becomes HOLD.
SYN
Illegal
After executing the TARM SGL command followed by the
occurrence of the LEVEL event,
sample event until the specified number of readings are
completed. The trigger arm event then becomes HOLD.
Illegal
After executing the TARM SGL command followed by the power
line voltage crossing zero volts, one reading is taken per sample
event until the specified number of readings are completed. The
trigger arm event then becomes HOLD.
Illegal
Any
Illegal
After executing the TARM SGL command, followed by the
controller requesting data
SYN
the first reading is taken. One reading is then taken per SYN
event until the specified number of readings are completed.
The trigger arm event then becomes HOLD.
After executing the TARM SGL command, followed by the
controller requesting data,
event until the specified number of readings are completed.
The trigger arm event then becomes HOLD.
[a]
one reading is taken per
[b]
, which satisfies both SYN events,
[b]
one reading is taken per sample
Keysight 3458A User's Guide
[c]
[c]

Publicité