– The multimeter sets bit 0 in the status register after executing a stored
– From the front panel, you can view all stored subprogram names by accessing
– Related commands: COMPRESS, CONT, DELSUB, PAUSE, SCRATCH, SUB,
Examples
OUTPUT 722;"CALL DCCUR2" !EXECUTES SUBPROGRAM NAMED "DCCUR2"
CALNUM?
Calibration number query. Returns an integer indicating the number of times the
multimeter has been calibrated.
Syntax
CALNUM?
Remarks
– The calibration number is incremented by 1 whenever the multimeter is
– The calibration number is stored in cal-protected memory and is not lost when
– The multimeter was calibrated before it left the factory. When you receive the
– Related commands: CAL, CALSTR, SCAL
Example
10 OUTPUT 722;"CALNUM?" !READS CALIBRATION NUMBER
20 ENTER 722;A!ENTERS RESPONSE INTO COMPUTER
30 PRINT A!PRINTS RESPONSE
40 END
Keysight 3458A User's Guide
subprogram.
the CALL command and pressing the up or down arrow key. Once you have
found the correct subprogram, press the Enter key to execute the subprogram.
SUBEND
calibrated. If autocal is secured, the calibration number is also incremented by
1 whenever an autocal is performed; if unsecured, autocal does not affect the
calibration number.
power is removed.
multimeter, read the calibration number to determine its initial value.
Command Reference
6
245