Command Reference
Power-on control = last programmed value.
Default control = ONCE.
Remarks
– The multimeter stores the control parameter in continuous memory (the
– Query command: The BEEP? query command returns the present beeper
– Related commands: TONE
Example
OUTPUT 722;"BEEP OFF" !DISABLES THE BEEPER
CAL
This is a calibration command. Refer to the 3458A Calibration Manual for details.
CALL
Call subprogram. Executes a previously stored subprogram.
Syntax
CALL [name]
name
Subprogram name. A subprogram name may contain up to 10 characters. The
name can be alpha, alphanumeric, or an integer in the range of 0 to 127. Refer to
the
Power-on name = none.
Default name = 0.
Remarks
– Subprograms are created with the SUB command.
244
parameter is not lost when power is removed).
mode. Refer to
Query commands
information.
SUB
command for details.
near the front of this chapter for more
Keysight 3458A User's Guide