Télécharger Imprimer la page

Keysight Technologies 3458A Guide D'utilisation page 253

Multimètre

Publicité

Command Reference
6
– Related commands: NRDGS, SWEEP, TIMER, TRIG
Examples
OUTPUT 722;"DELAY 5" !INSERTS A 5 SECOND DELAY
OUTPUT 722;"DELAY -1" !RETURNS TO AUTOMATIC (DEFAULT) DELAY
DELSUB
Delete subprogram. Removes a single subprogram from memory.
Syntax
DELSUB name
name
Subprogram name. A subprogram name may contain up to 10 characters. The
name can be alpha, alphanumeric, or an integer in the range of 0 to 127. Refer to
the
SUB
command for details.
Power-on name = none.
Default name = none; parameter required.
Remarks
– When a subprogram is deleted, the memory used to store it is freed and may
be used to store a new subprogram (see the
SUB
command).
– To delete all subprograms at once, use the SCRATCH command.
– Related commands: COMPRESS, SCRATCH, SUB
Example
OUTPUT 722;"DELSUB TEST12" !DELETES SUBPROGRAM TEST12
Keysight 3458A User's Guide
253

Publicité

loading