90 PRINT A !PRINTS STANDARD DEVIATION
100 END
RMEM
Recall memory. Reads and returns the value of a reading or group of readings
stored in reading memory. RMEM leaves stored readings intact (not cleared from
memory).
Syntax
RMEM [first][,count][,record]
first
Designates the beginning reading.
Power-on first = none.
Default first = 1.
count
Designates the number of readings to be recalled, starting with first.
Power-on count = none.
Default count = 1.
record
Designates the record from which to recall readings. Records correspond to the
number of readings specified by the NRDGS command. For example, if NRDGS
specifies three readings per trigger, each record will contain three readings.
Power-on record = none.
Default record = 1.
Remarks
– The RMEM command automatically shuts off reading memory (MEM OFF). This
– The multimeter assigns a number to each reading in reading memory. The
Keysight 3458A User's Guide
means all previously stored readings remain intact and new readings are not
stored. You can re-enable reading memory without destroying any stored
readings using the MEM CONT command.
most recent reading is assigned the lowest number (1) and the oldest reading
Command Reference
6