Télécharger Imprimer la page

Keysight Technologies 3458A Guide D'utilisation page 240

Multimètre

Publicité

6
Command Reference
– Related commands: FUNC, RANGE
Example
OUTPUT 722;"ARANGE OFF" !DISABLES AUTORANGE
AUXERR?
Auxiliary error. When a hardware error is detected. the multimeter sets a bit in
the auxiliary error register. The AUXERR? command returns a number
representing the decimal-weighted sum of all set bits.The register is then cleared.
Syntax
AUXERR?
Auxiliary error conditions
The auxiliary error conditions and their weighted values are:
Weighted
value
1
2
4
8
16
32
64
128
256
512
1024
2048
4096
240
Bit
Description
number
0
Slave processor not responding
1
DTACK failure
2
Slave processor self-test failure
3
Isolator test failure
4
A/D converter convergence failure
5
Calibration value out of range
6
GPIB chip failure
7
UART failure
8
Timer failure
9
Internal overload
10
ROM checksum failure, low-order byte
11
ROM checksum failure, high-order byte
12
Nonvolatile RAM failure
Keysight 3458A User's Guide