Command Reference
program places 15 readings (3 readings per trigger event, 5 times) into reading
memory.
Unless the input buffer is enabled, line 60 causes the GPIB bus to be held until all
measurement cycles are complete. If you want to regain control of the bus
immediately, suppress the cr lf by replacing line 60 with:
60 OUTPUT 722 USING "#,K"TARM SGL, 5;"
In the above line, the # image specifier suppresses the cr lf. The K image specifier
suppresses trailing or leading spaces and outputs the command in free-field
format. Notice the semicolon following the TARM SGL.5. This indicates the end of
the command to the multimeter and must be present when you suppress cr lf.
TBUFF
Trigger buffer. Enables or disables the multimeter's external trigger buffer.
Syntax
TBUFF [control]
control
The control parameter choices are:
control
parameter
OFF
ON
Power-on control = OFF.
Default control = OFF.
Remarks
– Setting TBUFF to ON corrects for a TRIGGER TOO FAST error that can occur
368
Numeric
query equiv.
0
1
when using an external EXT trigger arm, trigger, or sample event. With TBUFF
OFF, any external trigger occurring during a reading generates the TRIGGER
Description
Disables the trigger buffer which enables the TRIGGER TOO FAST
error
Enables and clears the trigger buffer which disables the TRIGGER
TOO FAST error
Keysight 3458A User's Guide