– If the multimeter's keyboard is disabled by both LOCAL LOCKOUT and the
Examples
10 REMOTE 722 !SETS DEVICE AT ADDRESS 22 TO REMOTE STATE
20 LOCAL LOCKOUT 7 !SENDS LOCAL LOCKOUT (LLO) TO ALL
30 END !DEVICES ON THE BUS
REMOTE
Sets the GPIB REN line true.
Syntax
REMOTE 7
REMOTE 722
Remarks
– The REMOTE 722 command places the multimeter in the remote state. The
– In most cases. you will only need the REMOTE command after using the
Examples
REMOTE 7 !SETS GPIB REN LINE TRUE
The above line does not, by itself, place the multimeter in the remote state. The
multimeter will only go into the remote state when it receives its listen address
(e.g., sending OUTPUT 722;"BEEP").
Keysight 3458A User's Guide
the keyboard but a subsequent remote command disables it. Sending the
LOCAL 7 command, however, enables the LOCAL key and keeps it enabled
even after a subsequent remote message.
LOCK command, you must clear both to regain control of the keyboard. LOCAL
LOCKOUT is cleared with the LOCAL command. LOCK is cleared by setting
LOCK to OFF.
REMOTE 7 command, does not. by itself. place the multimeter in the remote
state. After sending the REMOTE 7 command, the multimeter will only go into
the remote state when it receives its listen address.
LOCAL command. REMOTE is independent of any other GPIB activity and is
sent on a single bus line called REN. Most controllers set the REN line true
when power is applied or when reset.
GPIB Commands
415