3
Configuring for Measurements
Line 20 presets the multimeter, which suspends triggering. Line 30 clears the
status register. Line 40 instructs the controller to go to line 90 should an interrupt
occur. Line 50 enables SRQ interrupts on the GPIB interface. Line 60 enables the
hi/lo limit, power-on, and error bits to assert SRQ. Line 60 also enables the
real-time pass/fail math operation with the values of –5 for the low limit and +5
for the hi limit. Line 70 enables automatic triggering. Line 80 causes the controller
to wait for an interrupt. Lines 90 through 130 read the status register and print
which condition(s) caused the interrupt.