Les mesures
50 Num_readings=5000
60 ASSIGN @Dvm TO 722
70 ASSIGN @Int_rdgs TO BUFFER Int_rdgs(*) ! AFFECTATION DU NOM
75
80 OUTPUT @Dvm;"PRESET FAST;OFORMAT SINT;MFORMAT SINT;MEM FIFO;APER
l.4E-6;NRDGS ";Num_readings ! ARMEMENT SYN, DECLENCHEMENT AUTO,
85 ! LECTURES RAPIDES, FORMAT MEMOIRE/SORTIE ENTIER SIMPLE, 1,4 µs
87 ! TEMPS D'INTEGRATION, 5000 LECTURES/DECLENCHEMENT
90 OUTPUT @Dvm;"TARM SGL"
100 TRANSFER @Dvrn TO @Int_rdgs;WAIT ! TRANSFERT DES LECTURES DANS LE
105 ! TABLEAU DES ENTIERS; PAS DE CONVERSION DE DONNEES REQUISES
107 ! PUISQUE LE FORMAT INTEGER DE L'ORDINATEUR EST IDENTIQUE A SINT
109 ! (MAIS TABLEAUX DES ENTIERS REQUIS)
110 OUTPUT @Dvm; "ISCALE'?"
115
120 ENTER @Dvm;S
130 FOR I=l TO Num_readings
140 Rdgs(I)=Int_rdgs(I)
145 ! ENTIER AU FORMAT REEL(OBLIGATOIRE POUR EVITER UN
147 ! DEPASSEMENT ENTIER SUR LA LIGNE SUIVANTE)
150 R=ABS(Rdgs(I))
155
160 IF R>=32767 THEN PRINT "OVLD" ! SI SURCHARGE, IMPRESSION D'UN
165
170 Rdgs(I)=Rdgs(I)*S
175
180 Rdgs(I)=DROUND(Rdgs(I),4)! VALEUR ARRONDIE A 4 CHIFFRES
190 NEXT I
200 END
Détermination de la vitesse de lecture
Lorsque l'événement d'échantillonnage TIMER ou la commande SWEEP est
utilisé, la vitesse de lecture est simplement l'inverse de l'intervalle spécifié entre
les lectures (à condition que l'erreur TRIGGER TOO FAST ne se produise pas). Par
178
NOMBRE DE LECTURES = 5000
AFFECTATION DE L'ADRESSE DU MULTIMETRE
! D'ACCES DE LA MEMOIRE-TAMPON
! DECLENCHEMENT DES LECTURES
! INTERROGATION DU FACTEUR DECHELLE POUR
! LE FORMAT SINT
! LECTURE DU FACTEUR D'ECHELLE
! CONVERSION DE CHAQUE LECTURE DU FORMAT
! UTILISATION DE LA VALEUR ABSOLUE POUR
! VERIFIER SURCHARGE
! MESSAGE
! MULTIPLICATION DE LA LECTURE PAR LE
! FACTEUR D'ECHELLE