For ACI or ACDCI:
max._input
parameter
–1 or AUTO
0 to .120E-6
>120E-6 to 1.2E-3
>1.2E-3 to 12E-3
>12E-3 to 120E-3
>120E-3 to 1.2
For DSAC or DSDC:
max._input
parameter
0 to .012
>.012 to .120
>.120 to 1.2
>1.2 to 12
>12 to 120
>120 to 1E3
Power-on max._input = AUTO.
Default max._input = AUTO.
%
For most measurement functions, you specify the %_resolution as a percentage of
the max._input parameter. (Refer to the
showing how %_resolution affects frequency and period measurements;
%_resolution is ignored when the function parameter is DSAC, DSDC, SSAC, or
SSDC.)
For all functions except FREQ, PER, DSAC, DSDC, SSAC, and SSDC, the
multimeter multiplies %_resolution times max._input to determine the
measurement's resolution. For example, suppose you are measuring DC voltage,
your maximum expected input is 10 V, and you want 1 mV of resolution. To
determine %_resolution, use the equation:
Keysight 3458A User's Guide
Selects
range
Autorange
100 µA
1 mA
10 mA
100 mA
1 A
Selects
range
10 mV
100 mV
1 V
10 V
100 V
1000 V
_
resolution
For SSAC or SSDC:
max._input
Full scale
parameter
0 to .012
120 µA
>.012 to .120
1.2 mA
>.120 to 1.2
12 mA
>1.2 to 12
120 mA
>12 to 120
1.05 A
>120 to 1E3
Full scale
SINT
DINT
format
format
12 mV
50 mV
120 mV
500 mV
1.2 V
5.0 V
12 V
50 V
120 V
500 V
1050 V
1050 V
FREQ
Command Reference
Selects
range
10 mV
100 mV
1 V
10 V
100 V
1000 V
and
PER
commands for tables
6
Full scale
12 mV
120 mV
1.2 V
12 V
120 V
1050 V
273