The GPIB address
You can change the multimeter's GPIB address using the ADDRESS command.
Refer to
change the GPIB address. The multimeter leaves the factory with the address set
to decimal 22. The corresponding ASCII code is a listen address of 6 and a talk
address of V.
The examples in this manual are intended for Hewlett-Packard Series 200\300
NOTE
computers using the BASIC language. They assume a GPIB interface select code
of 7 and a device address of 22 resulting in a combined GPIB address of 722.
Mounting the multimeter
The multimeter comes equipped with four feet, which allow it to be used as a
bench instrument. It also has two tilt stands that allow you to elevate the front of
the multimeter. The multimeter can be mounted in a standard 19-inch rack using
the optional rack mount kits listed in
Installation verification
The following program verifies that the multimeter is operating and can
communicate with the controller over the GPIB bus.
10 PRINTER IS 1
20 OUTPUT 722;"ID?"
30 ENTER 722; IDENT$
40 PRINT IDENT$
50 END
If the multimeter has been correctly installed, the message
be printed on the designated system printer. If no message is printed, make sure
power is applied to the multimeter. Also check the GPIB connections, the
interface address setting, and the multimeter's address.
Keysight 3458A User's Guide
Changing the GPIB
address, in
Installation and Maintenance
Chapter
2, for a procedure on how to
Table
1-1.
will
Keysight 3458A
39
1