Command Reference
high
Specifies the highest expected frequency component of the input signal.
Power-on high_frequency = 20 MHz
Default high_frequency = 2 MHz
Remarks
– Refer to the
– For synchronous ACV or ACDCV (SETACV SYNC command), the bandwidth
– For frequency or period measurements with autorange enabled, the
– If you are unsure of the frequency content of the input signal, default the
– Query command: The ACBAND? query command returns two numbers
– Related commands: ACDCI, ACDCV, ACI, ACV, FREQ, FUNC, PER. SETACV
Example
OUTPUT 722;"ACBAND 500,1000" !SPECIFIES THAT THE INPUT SIGNAL
236
_
frequency
"Appendix A: Specifications"
reading rate specifications based on the bandwidth of the input signal.
parameters are used by the multimeter to calculate time-out values and
sampling parameters. When using level triggering (default mode), if the input
signal is removed during a reading and does not return within the time limits,
the measurement method changes to random so that the reading can be
completed. (After the reading, the measurement method returns to SYNC.) For
synchronous ACV or ACDCV, it is very important that the specified bandwidth
corresponds to the frequency content of the signal being measured.
bandwidth parameters are used to determine the amount of time needed for
autoranging. For these measurements, it is very important that the specified
bandwidth (especially low_frequency) corresponds to the frequency content of
the signal being measured.
ACBAND parameters.
separated by a comma. The first number is the currently specified
low_frequency, the second number is the high_frequency. Refer to
commands
near the front of this chapter for more information.
on page 409 for accuracy and
!IS BETWEEN 500 - 1000 Hz
Keysight 3458A User's Guide