Télécharger Imprimer la page

Keysight Technologies 3458A Guide D'utilisation page 254

Multimètre

Publicité

6
Command Reference
DIAGNOST
This is a service-related command. Refer to the 3458A Service Manual for details
DISP
Display. Enables or disables the multimeter's display, and may also be used to
send a message to the display or to clear the display.
Syntax
DISP [control] [,message]
control
The control parameter choices are:
control
parameter
OFF
ON
MSG
CLR
Power-on control = ON.
Default control = ON.
message
The message parameter is the message to be displayed. The message may
contain spaces, numerals, lower or upper case letters, and any of the following
characters:
! # $ % & ' ( ) ^ \ / @ ; : [ ] , . + - = * < > ? _
254
Numeric
Description
query equiv.
Displays message if included (if no message, dashes are
displayed); inactivates all annunciators except ERR; readings are
0
no longer displayed and the display is not updated except to
service front panel keystrokes and query commands.
1
Normal (power-on mode) display operation
2
Displays message, annunciators activated
3
Clears the display
Keysight 3458A User's Guide