– Executing the RESET command sets TBUFF to OFF.
– Query command: The TBUFF? query command returns the present trigger
– Related commands: EXTOUT, NRDGS, TRIG
Example
OUTPUT 722;"TBUFF ON" !DISABLES THE TRIGGER TOO FAST ERROR
TEMP?
Temperature query. Returns the multimeter's internal temperature in degrees
Centigrade.
Syntax
TEMP?
Remarks
– Monitoring the multimeter's temperature is helpful to determine when to
– Related commands: ACAL, CAL, CALSTR
Example
10 OUTPUT 722; "TEMP?" !READ TEMPERATURE
20 ENTER 722; A !ENTER RESULT
30 PRINT A !PRINT RESULT
40 END
Keysight 3458A User's Guide
TOO FAST error and the trigger(s) are ignored. With TBUFF ON, the first
external trigger occurring during a reading is stored and no error is generated
by this or any successive triggers. After the reading is complete, the stored
trigger satisfies the EXT event if the multimeter is so programmed.
buffering mode. Refer to
more information.
perform autocalibration.
Query commands
near the front of this chapter for
Command Reference
6
369