Making Measurements
Event combinations
You can specify many combinations of the trigger arm, trigger, and sample events
to suit your application.
events and describes the resultant triggering sequence for each.
Table 4-2
Trigger
136
Event combinations
Trigger
arm
Sample event Description
event
event
AUTO
AUTO
AUTO, EXT,
AUTO
EXT
TIMER, LINE,
LEVEL
AUTO
EXT
AUTO, EXT,
AUTO
LEVEL
TIMER, LEVEL
AUTO
LEVEL
SYN, LINE
AUTO, EXT,
AUTO
LINE
TIMER, LINE
AUTO
LINE
SYN, LEVEL
AUTO
SGL
AUTO
SYN
AUTO, EXT,
AUTO
SYN
LEVEL,
LINE, TIMER
Table 4-2
shows, all possible combinations of these
One reading is taken per sample event (if the sample event is
Any
AUTO, readings are taken continuously).
After a negative edge transition on the Ext Trig input, one reading
is taken per sample event until the specified number of readings
are completed.
SYN
Illegal
After the LEVEL event
event, until the specified number of readings are completed.
Illegal
After the power line voltage crosses zero volts, one reading is
taken per sample event until the specified number of readings
are completed.
Illegal
After executing the TRIG SGL command, one reading is taken per
sample event until the specified number of readings are
Any
completed. The trigger event then becomes HOLD. When using
the SYN sample event, the input buffer must be enabled or you
must suppress cr lf when sending the TRIG SGL command.
After the controller requests data,
satisfied and the first reading is taken. One reading is then taken
SYN
per SYN event until the specified number of readings are
completed.
After the controller requests data,
sample event until the specified number of readings are
completed.
[a]
occurs, one reading is taken per sample
[b]
both SYN events are
[b]
one reading is taken per
Keysight 3458A User's Guide