410 OUTPUT @Scan;"DMODE 1,1,0,1"! Setup for external increment
420 !
430 !
440 ! -------- Measurement Setup --------
450 !
460 OUTPUT @Dmm;"PRESET;TARM HOLD"!Sets the dmm in its normal PRESET
461 !
462 OUTPUT @Dmm:"MFORMAT DREAL"! Stores the data in memory in IEEE
463 !
470 OUTPUT @Dmm;"TRIG EXT"!Sets the dmm to trigger externally
480 OUTPUT @Dmm;"APER 20E-6"!Sets the integrator aperture to 20 µs
490 OUTPUT @Dmm;"TBUFF ON"! Sets-up the trigger buffer
500 !
510 OUTPUT @Dmm;"DISP OFF !
the dmm
520 OUTPUT @DMM;"DELAY 0"!
530 !
540 !
550 ! --------- Measurement List -------------
560 !
570 OUTPUT @Dmm;"SUB 1"! Start of the dmm program
580 OUTPUT @Dmm;"MEM FIFO"!
590 OUTPUT @Dmm;"DCV 10" !
600 OUTPUT @Dmm;"TARM SGL"! (1) Initiates the measurement sequence once
610 !
620 !
630 OUTPUT @Dmm;"TARM SGL"!(2)
640 OUTPUT @Dmm;"TARM SGL"!(3)
Keysight 3458A User's Guide
and channel close on the scanner.
and holds the trigger arm
double-real format
does not occur
and measurement start to 0 s
max
the
TRIG EXT is satisfied and stops after just
one trigger event occurs.
Optimizing Throughout and Reading Rate
Turns off the front panel display on
Sets the time between trigger event
Sets memory to first-in, first-out
Sets the dmm to dcV function and 10 volts
Repeats the sequence again.