Description
Block calibrators
Ohmic measurement
7.1
Cleaning
Cleaning the enclosure
● Clean the outside of the enclosure with the inscriptions and the display window using a cloth moistened with water or a
mild detergent.
● Do not use any aggressive cleansing agents or solvents, e.g. acetone. Plastic parts or the painted surface could be
damaged. The inscriptions could become unreadable.
WARNING
Electrostatic charge
Danger of explosion in hazardous areas if electrostatic charges develop, for example, when cleaning plastic surfaces with a
dry cloth.
Prevent electrostatic charging in hazardous areas.
7.2

Return procedure

Enclose the bill of lading, return document and decontamination certificate in a clear plastic pouch and attach it firmly to the
outside of the packaging.
Required forms
● Delivery note
● Return goods delivery note (http://www.siemens.com/processinstrumentation/returngoodsnote)
with the following information:
Product (item description)
Number of returned devices/replacement parts
Reason for returning the item(s)
● Decontamination declaration (http://www.siemens.com/sc/declarationofdecontamination)
22
Recalibration procedure
Loosen recessed grip screw(s) to remove the RTD connector or
housing and unscrew the measuring insert from the pipe collar.
Use calibrator sleeves that have been adapted to the shape of the
RTD unit only.
Insert must have a borehole of ∅6.00 mm (0.24") H7, depth = 8 mm
(0.31").
Do not exceed 100 °C (212 °F) at locking plug [80 °C (176 °F) when
using a temperature transmitter].
Use block calibrator with dual-zone-technology with internal
reference sensor only.
Observe the adjustment time specified by the manufacturer when
heating the calibrator.
1
2
3
4
1
2
SITRANS TSinsert/TS100/TS200/TS300/TS500 (7MC71.. 7MC72.. 7MC80.. 7MC75.. 7MC65.. 7MC55..)
Apply heat sink compound to the RTD unit before inserting it
in the calibrator sleeve.
Check the electrical connector (cable end) as indicated by
the nameplate.
After inserting the RTD unit, wait about 5 minutes for the
temperature to settle.
Compare the temperature of the calibrator with the RTD
temperature and adjust if necessary.
Take into account any line resistance.
Apply heat sink compound to the RTD plug-in unit.
A5E03920348-AB, 03/2016