146
BTA?
Reads the 64 bit integer value which can be written by the fieldbus master
Syntax :
BTA?(x)
Parameter:
none
Response:
q1(y): current value written by fieldbus master in hex
e.g. 0xab12
STF
Syntax:
STFp1 (x)
Parameter:
p1 = 0:default, ticks (increasing counter) as a 64 bit integer
value
p1 = 1: system time as two 32 bit integers, nanoseconds (1
nd
seconds (2
4 bytes)
p1 = 2:system time as two 32 bit integers, microseconds and seconds
p1 = 3:system time as two 32 bit integers, 2^-32 seconds and seconds
The sum of the seconds and its fractions is the time passed since 01.01.1970.
The system time can be derived from the NTP-time.
The accuracy is not 100% predictable.
Note:
Response depends on SRB command.
STF?
Reads the time format currently in use
Syntax :
STF?(x)
Parameter:
none
Response:
q1(y): current setting of the timeformat
Bus To Application Query
STF Set Time Format
Defines the content and format of the time channels (MCS
17,18,19)
STF? Set TimeFormat Query
Jeu de commandes
st
4 bytes) and
A3257−2.0 en/de/fr